Notice: Function _load_textdomain_just_in_time was called incorrectly. Translation loading for the updraftplus domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /home/kuykyqzrwdjm/public_html/wp-includes/functions.php on line 6114

Notice: Function _load_textdomain_just_in_time was called incorrectly. Translation loading for the wordpress-seo domain was triggered too early. This is usually an indicator for some code in the plugin or theme running too early. Translations should be loaded at the init action or later. Please see Debugging in WordPress for more information. (This message was added in version 6.7.0.) in /home/kuykyqzrwdjm/public_html/wp-includes/functions.php on line 6114
CPG Automation - CPG Automation

3rd party lab comparison of Integra Boost(tm) 62% RH and BOVEDA 62% RH 2 way control humectant

Bonner Analytical Testing Company | July 7, 2022

For product information please visit www.desiccarecanada.com

Report from January 26, 2017

Bonner Analytical Testing Company was retained by Desiccare Inc. to evaluate off gassing characteristics of their product Integra Boost 62% RH and a competitors product BOVEDA 62% RH 2 Way Control Humectant. The experiments were conducted January 25, 2017.


Experimental:
One Boveda 62% RH sachet and one Integra Boost 62% RH sachet was removed from its clear packaging and placed in separate five liter foil faced gas sampling bag equipped with a septum sampling port. The bags were then sealed. A third Control bag was treated in a similar manner


Two hundred milliliters of outside air was introduced to each bag.

The bags were allowed to sit at room temperature for 6 hours prior to analysis.


The analytical procedure involved removing an aliquant of air (5.0cc) from each of the bags, and injecting the samples into a gas chromatograph equipped with an Ocean Instruments Eclipse purge and trap introductory system connected to an Agilent 6890 gas chromatograph
with an Agilent 5973 mass selective detector.


The method quantitation limit (MQL) was estimated to be 1.0 ug/l.


Bonner Analytical Testing Co.
2703 Oak Grove Road, Hattiesburg, MS 39402
Phone No.: 601-264-9965 Fax No.: 601-268-7084
www.batco.com


Results:
Sample ID Analyte Concentration (ug/L) Duplicate
Control Acetone ND ND
Integra Boost 62% RH Acetone ND ND
Boveda 62% RH Acetone 9.8 9.35
ND = Non-Detect


Conclusion:
Boveda 62% RH 2-Way Humidity Control Humectant produces a significant amount of Acetone as a byproduct compared to Integra Boost 62% RH; which was Non-Detect (<1.0ug/l). Additionally, the Boveda product continued to off gas giving a concentration of 13.11 ug/l at 18